Reference change value (RCV)¶
What it answers. Two results of one marker differ. Is the difference larger than the normal day-to-day variation of the person and of the lab method? The RCV is the smallest change, in percent, that is unlikely (5% two-sided) to come from that variation alone.
Formula.
CVi is the within-person biological variation and CVa is the analytical variation of the method, both in percent. 1.96 is the two-sided 95% z value. √2 is there because both results carry the variation.
Source of the numbers. Every CV comes from the European Biological Variation Study (EuBIVAS). The
values are the ones in Table 2 of the summary report: Carobene A et al. The European Biological
Variation Study (EuBIVAS): a summary report. Clin Chem Lab Med 2022;60:505-517.
https://doi.org/10.1515/cclm-2021-0370. Each rule also names the primary EuBIVAS paper for its
marker in primary.
| Marker | Rule | CVi % | CVa % | RCV % | Primary paper |
|---|---|---|---|---|---|
| LDL cholesterol | rcv.ldl_c |
8.46 | 1.77 | 24.0 | https://doi.org/10.1373/clinchem.2018.288415 |
| hs-CRP | rcv.crp_hs |
29.4 | 11.1 | 87.1 | https://doi.org/10.1373/clinchem.2019.304618 |
| Creatinine (enzymatic) | rcv.creatinine |
4.4 | 1.1 | 12.6 | https://doi.org/10.1373/clinchem.2017.275115 |
| ALT | rcv.alt |
9.3 | 6.7 | 31.8 | https://doi.org/10.1373/clinchem.2016.269811 |
| TSH | rcv.tsh |
18.9 | 1.46 | 52.5 | https://doi.org/10.1515/cclm-2020-1885 |
Notes on the table:
- TSH. The summary table gives CVi 18.9%. The abstract of the primary paper gives 17.7%. We use the summary table, the same source as every other row.
- Creatinine. The values are for the enzymatic method. The alkaline picrate (Jaffe) method gave CVi 4.7% and CVa 4.4%. A Jaffe result and an enzymatic result are not compared, see below. The LOINC codes we ship do not tell the two methods apart, so a creatinine result with no method recorded is judged with the enzymatic values. An install that knows its lab uses Jaffe can override the rule.
- HbA1c and ferritin have no RCV in pack 0.2.0. We found no EuBIVAS estimate for them that we may
ship. An install can add a rule in its override directory (
OWNHEALTH_RULES_OVERRIDE_DIR). The pack version then ends in+local.<sha8>, so the change is visible in every answer.
Only like with like. A change is judged only between two results with the same method and the same lab (the provider name of the source document). The method is the one stored with the result, else the one the shipped LOINC table gives its code (calculated or direct LDL, the eGFR equation: CKD-EPI, CKD-EPI 2021 or MDRD). Two codes that differ only in the unit name the same method: creatinine in mg/dL (LOINC 2160-0) and in µmol/L (14682-9) compare through their canonical value. The comparison ignores case and extra spaces. A missing lab name or method counts as one more value: two results that both have no lab name count as the same lab, even when they came from two labs, and two results that both have no method count as the same method. A result with a lab name and one without differ. When the last two results differ in method or lab:
- the rule pack raises no RCV flag;
- the REST trend gives
exceeds_rcv: null,method_change: trueand a sentence inlimitations; - the MCP tool
get_trendmarks the point withmethod_change: trueand adds a limitation.
A trend rule also uses only the points with the same method and lab as the last point.
A rule must cite its paper. The pack refuses to load an rcv or baseline rule whose cite is
not a DOI link (https://doi.org/...).
Where to read it.
- REST:
GET /api/profiles/{p}/trend?code=...returnsstats.rcv_pct,stats.exceeds_rcv,stats.cv_i,stats.cv_a,stats.rcv_citeandstats.method_change, andmethod_changeon each point. - MCP:
get_trendreturnsmethod_changeon a raw point that changed method or lab. - UI: the "Is the last change real?" card of a biomarker page shows the RCV, the two CVs and a link to their source.
Code. server/crates/ownhealth-rules/src/engine.rs (rcv_percent, rcv, method_key). Tests:
server/crates/ownhealth-rules/tests/rule_pack_020.rs.